Minutes W44

Spoke CM10-2

  • CTS test started on last Friday: both cavities reached resonance frequency with no issue (motor changed wrt previous test). V 
    • Some parameters: sensitivity 0.177 Hz/step, 1 motor shaft turn corresponds to 256 steps (gear ratio). Stepper motor 200 step/rev.
    • previous high power test done NOT at resonat frequency, power source frequency shifted to do the test (but piezo test not meaningful, if done outside resonance)
  • Issue in water flow in coupler line
  • At FREIA 3 methods are used to evaluate QL" 
    • VNA (always lowest result)
    • Decay method
    • LFD measurement:  cav1 ~ 300 Hz, cav2 ~280 Hz. Unipolar piezo can cover all the LFD range (580 Hz tot) V
  • Eacc given using in-kind Qt, but results using Tom Powers method (reflected power, energy stored method) are always more optimistic
  • Field emission measured for both cavities (~500 uSv/h onset 8 MV/m). NB: detector very close to module, ~30 cm away.
  • HL static always lower than dynamic 
    • coupler cooling line issue: absence of flow in cooling line X.
    • Necessary to drain water from cooling line to start it again and proceed with the test.
  • Warm-up: issue with level sensor (LT01) X 
    • OLD sensor, not changed by ESS-IFJ team
  • Some small He signal during warm-up: is it a real issue or common to all modules (open question)
  • Next module to be tested CM09: DK installation on Nov 3rd

Planned delivery of CM10 to ESS: w46